Rigel 62353



The Rigel 62353 is the latest hand-held safety Analyzer from Rigel Medical, able to test to the new standard for in-service and after repair testing of medical electronic devices, the IEC 62353 (2007).


Send Inquiry Call 877.254.7086

Overview

The Rigel 62353 is the latest hand-held safety Analyzer from Rigel Medical, able to test to the new standard for in-service and after repair testing of medical electronic devices, the IEC 62353 (2007).

This cost effective safety Analyzer meets all the IEC 62353 requirements and beyond. From record keeping to comparison of data when used in combination with Med-eBase, the Rigel 62353 is a truly versatile instrument for every service engineer demanding portable test equipment.

Its large internal memory and connectivity to Med-eBase, comprehensive database software, ensures fast and easy download of test results, managing your asset database, creation of test sequences and producing test certificates.

Specification

Power Measurement:
Method: VA rating
Range: 0.1KVA – 4KVA
Accuracy: ±10% + 2 counts
Mains Outlet Test:
Input Voltage Range: 0-300V AC, max current 16A
Measures: L – E, N – E & L – N
Accuracy: ± 5% of reading + 2 counts
IEC Mains Lead Test:
Test Duration: 2s
Test: Continuity of all conductors, Earth bond, Insulation and Polarity
General Specifications:
Mains Power: 230 VAC ±10%, 50Hz +/- 1Hz 120 VAC ±10%, 60Hz +/- 1Hz (USA model)
Battery: 6 x 1.5V Alkaline AA
Weights: 1.6 kg including batteries
Size: (L x W x D) 270 x 110 x 75 mm / 10.5 x 4 x 3"
Operating Conditions: 0' - 40'C, 0-90% RH - NC
Storage Environment: -15' - +60'C
Environmental Protection: IP 40
Part Numbers:
331A930 UK (13A)
331A931 Australian (10A)
331A932 Schuko (16A)
331A933 USA (15A-120V-60Hz)
331A934 Danish (13A)
331A935 French (16A)
331A936 German (16A)


Request info