Rigel 288



The Rigel 288 has become the most recognised electrical safety Analyzer on the market today. Its unique design, patented earthbond technology and use of internal memory make the Rigel 288 a truly valuable tool for every biomedical test professional.


Send Inquiry Call 877.254.7086

Overview

The Rigel 288 has become the most recognised electrical safety Analyzer on the market today. Its unique design, patented earthbond technology and use of internal memory make the Rigel 288 a truly valuable tool for every biomedical test professional.

Combining automatic and manual test sequences, data logging and direct printing facilities, the Rigel 288 is the most compact electrical safety Analyzer of its kind.

Versatility comes for conformance with several international standards such as IEC / 62353, 60601-1, NFPA-99 and many local variants.

Bluetooth connectivity to a range of accessories and Med-eBase, comprehensive database software, ensures fast and easy download of test results, managing your asset database, creation of test sequences and producing test certificates.

Specification

Earth Continuity:
Method: 2 wire technique, using 'zero' lead function.
Test Current: >+200mA, -200mA DC into 2 ohms
Max Test Voltage: 4-24V rms o/c (6V for IEC 60601 tests)
Measuring Range
(low range):
0.001 – 0.999 ohms @ 0.001 ohms resolution
Measuring Range
(mid range):
1.00 – 9.99 ohms /@ 0.01 ohms resolution
Measuring Range
(high range):
10.0 – 19.9 ohms @ 0.1 ohms resolution
Accuracy: ± 3% of reading + 10 m ohms
Insulation Resistance:
Measurement: EUT to Earth / Ground, EUT to AP, AP to Ground
Voltage: 250V DC, 500 V DC @1mA
Range (low range): 0.01Mohms - 20 Mohms
Accuracy (low range): ± 5% of reading +2 counts
Range (high range): 20Mohms – 100Mohms
Accuracy (high range): ±10% +2 counts
Resolution: 0.01Mohms
Direct Leakage Measurement:
Measuring Range: 4µA to 9999µA
Accuracy: ± 5% or reading +2 counts
Mains on A.P. voltage: F-type only @ 110% of mains
Measuring Device: As per IEC 60601-1 requirements
Measurement Type: Separate AC & DC for Patient (-Auxiliary)
True RMS for all remaining Leakage tests
Leakage to IEC 60601
Differential Leakage Measurement:
Measuring Range: 75µA to 9999µA
Accuracy: ±5% of reading + 5 counts
Measurement / display resolution: 1µA
Measurement Type: True RMS
Measuring Device: Similar frequency response characteristics to IEC 60601-1
Alternative Leakage Measurements:
Test Voltage: 250V at mains frequency
Test Current: 3.5 mA current limited
Measurement Range: 4µA to 9999µA
Measurement Resolution: 1µA
Measurement Accuracy: ±5% of reading + 2 counts
Measurement Type: True RMS
Measuring Device: As per IEC 60601-1
Power Measurement:
Method: VA rating
Range: 0.1KVA – 4KVA
Accuracy: ±10% + 2 counts
Mains Outlet Test:
Input voltage range: 0-300V AC, max current 16A
Measures: L – E, N – E & L – N
Accuracy: ± 5% of reading + 2 counts
IEC Mains Lead Test:
Test Duration: 2s
Test: Continuity of all conductors, Earth bond, Insulation and Polarity
General Specification:
Mains Power: 230 VAC ±10%, 50Hz +/- 1Hz 120 VAC ±10%, 60Hz +/- 1Hz (USA model)
Battery: 6 x 1.5V Alkaline AA
Weights: 1.6 kg including batteries
Size: (L x W x D) 270 x 110 x 75 mm / 10.5 x 4 x 3"
Operating Conditions: 0' - 40'C, 0-90% RH - NC
Storage Environment: -15' - +60'C
Environmental Protection: IP 40
Part Numbers:
331A910 UK (13A)
331A912 Australian (10A)
331A913 Schuko (16A)
331A914 USA (15A-120V-60Hz)
331A915 Danish (13A)
331A917 French (16A)
331A918 Czech / Polish (16A)
331A919 German (16A)


Request info