Rigel 277 Plus



The Rigel 277 Plus is a portable medical electrical safety analyzer combining IEC/EN 60601-1 compliance with additional test facilities for IEC/EN 61010 (Laboratory Equipment).


Send Inquiry Call 877.254.7086

Overview

The Rigel 277 Plus is a portable medical electrical safety Analyzer combining IEC/EN 60601-1 compliance with additional test facilities for IEC/EN 61010 (Laboratory Equipment) including Touch Leakage, Voltage Measurement and dedicated IEC 61010 Measuring Device (Body Model).

Customised test routines can be created to meet local standards and variants of safety test standards. The Rigel 277 Plus safety Analyzer performs manual, automatic and semi automatic test routines allowing full control of power up and power down sequence, thus reducing the overall test times.

The Rigel 277 Plus boasts a large full graphics display, internal memory, QWERTY keyboard and printer making this a completely stand alone medical safety Analyzer.

Specification

AC Earth Bond Test:
Test Current: 100mA, 1A, 10A, 25A
Current Source: Regulated
Open Test Voltage: 6V rms nominal
Range: 0.01ohms - 19.99ohms / 0.01ohms counts
Accuracy: ± 5% of reading, ± 2 counts
Measurement: 4-wire, floating earth
Insulation Test:
Test Voltage: 350 / 500V d.c. nominal
Auto Ranging: 100kohms - 20 Mohms (L) / 0.01Mohms counts
20Mohms – 100Mohms (H) / 0.01Mohms counts
Accuracy: ± 5% of reading, ± 2 counts (L)
±10% of reading, ± 2 counts (H)
Open Circuit Voltage: <750VDC
IEC 60601 Leakage Test:
Supply Voltage: Incoming mains (230/110VAC)
F-Type Open Circuit Voltage: 110% +/-20% of Mains Input Voltage
Maximum Load: 16A (13A for UK model)
Input Impedance: 1Kohm – as per 60601 MD
Frequency Response: as per IEC 60601-1 requirements
Range: 0.000mA – 9.999mA / 0.001 counts
Accuracy: ± 3% of reading, ± 4 counts
IEC 61010 Leakage Test:
Supply Voltage: Incoming mains (230/110VAC)
Maximum Load: 16A (13A for UK model)
Input Impedance: 2Kohms – as per 61010 MD
Frequency Response: DC – 2.5kHz (as per IEC 61010)
Range: 0.10mA – 3.50 mA / 0.01 counts
Accuracy: ± 3% of reading, ± 4 counts
Load Test:
Measured Load: 0.06 - 4kVA
Measured Voltage: Mains Supply +/-10%
IEC Lead Test:
Test: 40V a.c., 1mA nominal
Detects: Good, Open, Short, Reverse
General Specification:
Size: 400mm x 320mm x 160 mm
Weight: Appropriately 9 kg
Operating: 0ºC to 40ºC (non condensing)
Storage: -10ºC to 50ºC (non condensing)
Maximum R.H: 90%
Part Numbers:
350A910 UK / Ireland
350A913 Euro - Schuko
350A917 French
350A915 Danish
350A918 Czech / Polish
350A914 US (110V – 60Hz)
350A912 Australian


Request info